Model MoniTurb-C-S

The MoniTurb-C-S process turbidity sensor is an inline measuring system designed for the precise detection of turbidity in liquids. Using the 90° light scattering principle, it delivers reliable and highly accurate measurement results, even at very low particle concentrations. Its robust construction and low maintenance requirements make the system ideal for continuous monitoring and process control in industrial applications.

Common Applications and Industries

Chemical Industry
Petrochemical Industry
Brewing Industry
Beverage Industry
Food Industry
Filter Inlet
Filter Outlet
Quality Control
Well Water

The MoniTurb-C-S operates using the 90° side-scattered light principle (nephelometry). A light source in the NIR or full-spectrum range (approx. 380–1100 nm) projects a parallel beam through the process medium. Suspended particles scatter the light, and a silicon photodiode detector positioned at 90° measures the intensity of the scattered light, which is proportional to the particle concentration.

A ratio measurement of scattered light and transmitted light compensates for changes in product colour, fouling of the measuring windows and ageing of the light source, ensuring stable and reliable measurements. Depending on the application and calibration, the sensor accurately measures low to medium turbidity levels and can be calibrated in a wide range of units, including ppm, NTU and EBC.

The MoniTurb-C-S sensor is designed for low maintenance and long service life. Integrated compensation for product colour changes and fouling of the measuring windows ensures stable and reliable measurements during continuous operation. High-quality components, including a long-life NIR LED with a typical service life of up to 5 years, or an optional gas-filled measuring lamp with a service life of 1–2 years, provide dependable long-term performance.

The robust measuring cell is manufactured from stainless steel 1.4404 with sapphire measuring windows, allowing reliable operation under demanding process conditions at temperatures up to 140°C and pressures up to PN16 / ANSI Class 150. Flexible process connections, including DIN, ANSI and NPT, together with various material options, enable adaptation to a wide range of applications. An optional ATEX-certified Ex version allows safe operation in Zone 1 and Zone 2 hazardous areas.

SpecificationValue
Nominal SizesDN 25 – DN 150 / ½” – 6″
Process PressurePN 16 / ANSI Class 150
Process TemperatureMax. 140°C
Sensor MaterialStainless steel 1.4404
Window MaterialSapphire
Seal MaterialApplication-specific
Measuring RangeTypically 0–1 NTU, 0–300 NTU
Reproducibility±1%
Detector SystemSilicon photodiodes
Protection RatingIP65 / NEMA 4X
CleaningOptional cleaning probe / CIP compatible
Optional Ex ProtectionATEX Zone 1 / Zone 2

The MoniTurb-C-S operates using the 90° side-scattered light principle (nephelometry). A light source in the NIR or full-spectrum range (approx. 380–1100 nm) projects a parallel beam through the process medium. Suspended particles scatter the light, and a silicon photodiode detector positioned at 90° measures the intensity of the scattered light, which is proportional to the particle concentration.

A ratio measurement of scattered light and transmitted light compensates for changes in product colour, fouling of the measuring windows and ageing of the light source, ensuring stable and reliable measurements. Depending on the application and calibration, the sensor accurately measures low to medium turbidity levels and can be calibrated in a wide range of units, including ppm, NTU and EBC.

The MoniTurb-C-S sensor is designed for low maintenance and long service life. Integrated compensation for product colour changes and fouling of the measuring windows ensures stable and reliable measurements during continuous operation. High-quality components, including a long-life NIR LED with a typical service life of up to 5 years, or an optional gas-filled measuring lamp with a service life of 1–2 years, provide dependable long-term performance.

The robust measuring cell is manufactured from stainless steel 1.4404 with sapphire measuring windows, allowing reliable operation under demanding process conditions at temperatures up to 140°C and pressures up to PN16 / ANSI Class 150. Flexible process connections, including DIN, ANSI and NPT, together with various material options, enable adaptation to a wide range of applications. An optional ATEX-certified Ex version allows safe operation in Zone 1 and Zone 2 hazardous areas.

SpecificationValue
Nominal SizesDN 25 – DN 150 / ½” – 6″
Process PressurePN 16 / ANSI Class 150
Process TemperatureMax. 140°C
Sensor MaterialStainless steel 1.4404
Window MaterialSapphire
Seal MaterialApplication-specific
Measuring RangeTypically 0–1 NTU, 0–300 NTU
Reproducibility±1%
Detector SystemSilicon photodiodes
Protection RatingIP65 / NEMA 4X
CleaningOptional cleaning probe / CIP compatible
Optional Ex ProtectionATEX Zone 1 / Zone 2
SIGNAL AMPLIFIERS

Model C1

Description

The Model C1 Signal Amplifier is compatible with most optical sensors from the Monitek and Chemtronic product ranges. System programming and calibration are carried out via a PC or laptop using intuitive menu-driven software. Calibration can be performed with one or multiple reference samples. All configuration settings and calibration data can be securely stored on the PC’s hard drive for backup and future use.

  • Operating temperature: Up to 45 °C
  • Housing material: Stainless steel 1.4301
  • Measuring range: Sensor-specific (programmable)
  • Interface: USB
  • Software: Configuration and calibration via Windows PC
  • Reproducibility: ±1%
  • Protection class: IP65 / NEMA 4X
  • Power supply: 90–260 VAC, 50–60 Hz (optional: 24 VDC ±10%)
  • Analog output: 4–20 mA, 500 Ω (isolated)
  • Relay outputs: Maximum contact load 24 VDC / 1 A

Model Messenger (MSG)

Description

The Messenger (MSG) Signal Amplifier is compatible with all optical sensors from the Chemtronic and Monitek product series. It allows users to operate up to four single-channel sensors simultaneously. Independent operation of different sensors is also supported without any limitations. The measurement results from the individual sensors can be linked and processed in virtually any combination, enabling applications such as the straightforward implementation of dosing systems.

The system is configured and operated via a PC or laptop using intuitive menu-driven software. As an option, the Messenger is also available with an integrated panel PC for standalone operation.

  • Power Supply: 90–260 VAC, 50–60 Hz (optional: 24 VDC ±10%)
  • Power Consumption: Max. 50 VA
  • Relay Outputs: 4 freely programmable relays (48 V / 2 A)
  • Analog Outputs: Up to 4 × 0/4–20 mA, galvanically isolated
  • Communication Interfaces: RS-232C / RS-485 (Modbus RTU)
  • Digital Inputs: 4 × (optional), 5 V High
  • Reproducibility: ±1%
  • Operating Temperature: −10 °C to +50 °C
  • Housing: Stainless steel 1.4301, IP65 (NEMA 4X)
  • Optional Explosion Protection: ATEX Zone I / Zone II
CONTACT

Looking for the right solution?

Benefit from more than 50 years of experience, field-proven measurement technology and personal consultation. Together, we will find the right solution for your application. Visit our contact page and tell us about your project to receive a non-binding quotation.