Main Menu
This menu helps in navigating to various Messenger options.
Choose Main Menu
Calibration Menu
Main Menu > Calibration Menu
This menu is
used to correlate raw data collected from the sensors (photo detectors) corresponding to
a specific standard or sample concentration. Up to 8 different samples can be
read by the instrument. From this information the instrument can construct a lookup
table.
You have two calibration tables. Factory calibration table
and Process calibration table. Choose the desired calibration table, which needs to be
changed by selecting the respective button.
Factory calibration table can be modified / viewed using
FACTORY level Access permissions.
Choose Main Menu > Setup Inputs / Outputs Menu. You will see the following Options.
Com Port
Main Menu > Setup Inputs/Outputs > Set Com Port.
This menu is used to specify the Port settings. Speeds of up to 57K are allowed with an option to customize most of the standard options of RS-232/485 communications.
Choose COM Toggle Button to Specify the Com Port ( 1..4) that can be used for communication with the Messenger Device.
Choose Baud Toggle Button ( 4800 / 9600/ 19200/ 38400/ 57600 ) to Specify the Baud rate of Communication.Select the Modbus toggle button to switch around the transmission protocol of "Modbus normal" to "Modbus slow" (communication with older computers).
Analog Outputs
Main Menu > Setup Inputs/Outputs> Set Analog
Outputs.
The system has FOUR independent and assignable Analog 0/4 -20 mA loops, which can be connected to External Devices. Each output will operate from a user-defined RESULT equation. The sub-menus allow you to set all the necessary parameter values for Low and High limits and the required mode for the analog output.
Overrange
Analog Outputs can flag abnormal conditions or upset conditions that can be caused by Malfunction of Optics or by some material coating over the windows. We can assign a range of mv for normal operation. If the range exceeds, a fault is registered and an appropriate error message ( OPTICS FAULT ) will be displayed on the LCD and Measurements Screen.
How to configure Analog Outputs
Alarms
Main Menu > Setup Inputs/Outputs > Set Alarms
You can setup Alarm set conditions. Each Alarm can have a HYSTERESIS and a TIME DELAY (the condition must be met for at least this time before the alarm is reported). The alarm mode (LOW , HIGH , LOW LATCH , HIGH LATCH) can be selected per individual Alarm.
Alarms can be assigned to the following
signals
The different Alarm Types are
Low
Alarm is active when the
measured value falls below the specified LIMIT.Alarm is released
automatically
when the measuring value is above the specified
LIMIT.
Low Latch
Alarm is active when the measured value falls below the
specified LIMIT.Alarm should be released manually using RESET ALARMS menu
High.
Alarm is active when the measured value falls
above the specified LIMIT.Alarm is released automatically when the measuring
value is below the specified LIMIT.
High Latch
Alarm is active when the measured value falls
above the specified LIMIT.Alarm should be released manually using the RESET ALARMS menu
Choose on the Alarm (1-4) .
Specify the signal to which the Alarm needs to be assigned to, by Choosing on the Assigned Signal Toggle Button.
Choose Type Toggle Button to specify the type of Alarm.
Choose Limit text box to specify the mv. Choose Delay on Make/Break Text box to specify the delay in which the condition needs to be met at least before the alarm is reported.
Choose on Hysteresis to Specify the Hysteresis value.
The Hysteresis value will not have any effect for High and Low Latches.
In case any of the Alarms are not used, make sure that Type and Assigned Signal are "NOT ASSIGNED".
Set Cleaning Event
You can program up to TWO cleaning events.
A cleaning event is an automatic cycle that allows the optics to be CLEANED by the introduction of a cleaning solution to flow through the cell.
For each Cleaning Event, you can specify the Cleaning Frequency , Cycle Duration and Purge Time ( Time to allow the Sample to flow and purge out the cleaning solution). A Relay can be assigned to a Cleaning Event which will be active only during the Cycle Duration.
Cleaning Events can be activated in 2 ways,
Automatically or through Assigned Digital
Inputs .
Automatic
Choose Cleaning Event 1..2
Specify the Cleaning Frequency , Cycle Duration , and Purge Time. The Cleaning Event gets activated.
Once you save the settings.
Assign Digital Inputs
Choose Cleaning Event 1..2
Set Cleaning Frequency in Hours = 0
Specify the Cycle Duration and Purge Time. Save the settings. Activate the Relay through Assign Digital Inputs. The Cleaning Cycle gets activated ONLY ONCE.
Whenever you are trying to activate a Cleaning Event through Assign Digital inputs make sure that the Cleaning Frequency in Hours is 0. Failing so may cause malfunctioning of the Cleaning Event.
Set Access Code
You can set personal access code for Operator , Technician and Factory Levels preventing unauthorized personnel from altering key parameters that affect the operation of the unit.
Choose Operator text box. An Interface pops up Specify the Operator Password.
Choose Technician text box. An Interface pops up Specify the Technician Password.
Choose Factory text box. An Interface pops up Specify the Factory Password.
The selected Sensor number appears on the top. The data
present in the selected calibration table (Process/Factory) can be
viewed/edited using the following option.
Samples NOT used in Calibration process must
always contain a 0.
Process Calibration
Build Calibration
Repeat steps 2 to 7 for each
sample.
Make sure that the entered samples are evenly distributed over
your range.
For example at a range 0-100 g/L you can use samples of
approx. 0 gr/L, 30 gr/ L, 60 gr/L and 100 gr/L for calibration.
This menu is provided with diagnostic tools.
Choose Main menu > Diagnostic Menu,you will see a list of Diagnostic options.
Hardware Test
This menu is used for testing of hardware.
Lamp Control
You can view the current lamp's status and Lamp Intensity.
OK in Lamp Status indicates BOTH Lamps are
Working Correctly.
MAIN FAIL in Lamp Status indicates INTERNAL Lamp Failure.
EXT FAIL
in Lamp Status indicates EXTERNAL Lamp Failure.
BOTH FAIL
in Lamp
Status indicates INTERNAL and EXTERNAL Lamp Failures.
Choose Lamp Intensity text
box to specify the percentage of Intensity within the range 80% to 120%.
Choose Display Fault Toggle( INT LAMP / EXT LAMP / BOTH LAMPS ) button to explicitly generate Faults.
Choose Turn Lamp ON / Turn Lamp OFF Toggle Button to specify Whether Lamp needs to be Switched ON / OFF.
Choose External Lamp ON / External Lamp OFF Toggle button to specify whether External Lamp needs
to be ON/OFF. External Lamps are operated with an optionally available separate power pack.
Test Analog Outputs
Main Menu >Diagnostic Menu > Hardware Test >
Test Analog Outputs.
This Menu displays the mv of the FOUR Analog Outputs. A Fault in the Analog Output is displayed as "FAULT: 2mA" where 2mA is the current assigned during the fault condition. You can also test the Analog outputs using this option.
For Example:-
Set Analog Out1 to the following parameters
OUTPUT = Result1
Low = 0mv, High = 100mv
Mode = 4-20mA
OVERRANGE = CH1
Low mv = 0 , High mv = 150
Choose Analog OUT1
Enter the 50 Milli volts in the Analog Output text box and choose the "Update Outputs" key. The current observed at the Analog OUT1 should be 12mA.
Choose Restore Previous to restore the Analog Outputs to previous values before the entry of 50 Milli volts.
View Channel Output
Here you can view the outputs from all the 4 channels and Messenger Ambient Temp.
Raw Data
Raw Data is channel mv displayed irrelevant of ZERO DARK CURRENT.
Channel Signal
Channel signal is Channel mv displayed after subtracting ZERO DARK CURRENT.
Choose Show Raw Data / Show Channel Signal Toggle Button to switch between the display of Raw data and Channel signal.
Adjust Hardware Settings
Using this menu. You
can adjust the hardware settings.
Content to be Filled.
Zero Dark Current
Main Menu > Diagnostic Menu > Adjust Hardware Settings >
Zero Dark Current.
You can store the Zero Dark Current using this menu.
Choose "ZERO CHANNELS" to store the dark current. The lamps are turned OFF and the system will take a reading for ALL the channels and stores it as Dark current.
Choose "Restore Defaults Settings" to set the Zero Dark Current to 0.
Select Language
You can select the language which needs to be displayed. Currently four languages are available for selection.
Choose any of the language buttons to switch to that particular
language.
Set Access Level
You can enter the Access Code. Based on the Access code entered the Access Level changes.
Choose New Access Code text box. An Interface pops up Specify the new Access Code.
On entering incorrect access code the Access level will be defaulted to OPERATOR.
Choose the type of Fit to be performed on the Calibration data. There are 4 options available.
Choose Setup Sensor
Equations to view/edit the Sensor Equations.
The Setup Results are used for formatting the output that is to be displayed to the user, which is captured from different sensors
Choose Main Menu > Setup Results Menu. You will see different Options to format the output
Results NameTag
Choose the Result (1..4) Text box to specify the Nametags to be displayed for results captured in the Measurements Screen.
Example: - If the data captured in Result1 is Turbidity then the Nametag for Result1 can be given as "Turbidity".
Choose the Result ( 1..4) text box to specify the Result Equations.
Rules for Result Equations
Example : -
The Equation is constructed by "term1 [op] term2[op] term3 [op] term4 + offset" where each term can be a sensor or the logarithm of sensor and has a numerical factor(i.e. 3.67LogSEN4).The operators [op] can be +,-,*,/
Example:-
If you want to subtract the reading of the first sensor (Sensor 1) from the reading of the second sensor (Sensor2).
Main Menu > Setup Results Menu
>Setup Graphical Results.
You can format the Graphical display in Measurements screen. Check on Show Result checkboxes to select which of the results need to be displayed during the Graphical display.
Choose Range of Values to specify the Lower & Upper Limits of the Graph.
Choose Time Step in Minutes to specify the time duration with which the X-Axis Scale needs to be displayed.
Choose Show Color to select the color with which RESULT (1..4) need to be displayed in BAR / LINE Graph.
Time Step in seconds can be specified by entering decimal values Ex :- 0.1 , 0.2.
Main Menu > Setup Results
Menu> Setup Numerical Results.
You can format the Numerical display in Measurements screen. Check on relevant Show Result checkboxes to select which results need to be displayed during the Numerical display in Measurements screen.
Choose Font Size Toggle Button to specify the Font Size (Small / Medium/ Large ) with which the results need to be displayed.
Main Menu > Setup Results
Menu > Setup Data Logging.
You can specify the Time Interval (Sec) with which the
data needs to be logged in buffer for each result.
Choose Store Result1 ( 1..4) text box to specify the time in seconds.
Choose on Set Buffer Size to Specify the Buffer Size for each Result.
In order to default the values Choose Set Default Values (i.e. 60 Seconds).
Clear Process Calibration
Main Menu >Diagnostic
Menu > Clear/Reset Menu >Clear Process
Calibration.
Choose Sensor (1..4).
On Selection of Sensor, choose Copy From Factory Table key.
Factory calibration table contents of the Selected Sensor are copied to the Process calibration table.
On Selection of Sensor, choose Clear Table key. Process calibration table contents of Selected Sensor are defaulted to 0.
Clear Factory Calibration
Main Menu >Diagnostic Menu > Clear/Reset Menu
>Clear Factory Calibration.
Choose Sensor (1..4).
On Selection of Sensor when Clear Table key is entered. Factory calibration table contents of selected sensor are defaulted to 0.
Digital Inputs
Main Menu >Setup Inputs/Outputs>Assign Digital
Inputs.
The Digital Inputs are usually connected to External Devices Which may want to send a command to the Messenger Device. This Helps the Messenger Device to synchronize with other devices that are a part of the process being monitored.
Following are the commands that can be issued from the External devices to the Messenger Device.
Choose Input (1..4) to Assign the commands that are mentioned above to the Digital Inputs.
Choose Restore Default Settings to switch back to the Default parameters.
Relay Functions
Main Menu > Setup Inputs/Outputs > Set Relay
Functions.
This menu allows the user to assign the FUNCTION of the Relay.
The Functions available are :
Main Menu > Setup Results Menu >Setup Sensor
Equations.
Choose the Sensor (1..4) text box to specify the Sensor Equations.
Rules for Sensor Equations
Rules :
-
The Equation is constructed by "term1 [op] term2[op]
term3 [op] term4 + offset" where each term can be a channel or the
logarithm of channel and has a numerical factor(i.e. 3.67LogCH4).The
operators [op] can be +,-,*,/
Example
1:
Scatter Light
Turbidity:
Example
2:
Absorbance:
Example
3:
Dual Channel Absorbance
(Color):
Main Menu > Setup Results Menu > Set
Results Units.
Choose the Result (1..4) Text box to specify the units to be displayed for results captured in the Measurements Screen.
Example: - If the data captured in Result1 is
Turbidity then the unit of Result1 can be given as "NTU".
Typical Units are ppm, TEF, FTU, NTU, mg/l, g/l, %, APHA, etc..
Test Relays
This Menu displays the current status of the FOUR Relays
i.e. Signal the Relay has been assigned and whether it is switched
ON / OFF .
Choose the ON / OFF Toggle button to Switch ON / OFF the Relay.
Review/Reset Alarms
This menu displays the status of the 4 Alarms.
An ACTIVE in the status block with RED background indicates Alarm is ON.
In order to clear the Alarm, choose the Alarm with ACTIVE in the Status block. and choose the Clear ALARM key.
Start / End Menu
Select Language
Select Language in which you want to view the help in
Bubble Reject Threshold
Bubble Reject filter is used to smooth system response when bubbles are passing. and is based on a data averaging and acts upon a threshold limit set by the user in % change detection.
Choose Bubble Reject text boxes to specify the % of Threshold limit.
Choose Filter Delay text boxes to specify the delay in sec.
Main Menu > Setup Results Menu > Set
Results Format.
You can format the results displayed in Measurements screen.
Choose Number of Decimal Places for each Result to specify the precision with which the result needs to be displayed in Measurements Screen.
Choose Allow Values for each Result to specify Whether POS( Positive values) or ALL ( Positive and Negative values ) to be displayed in Measurements Screen.
Optimize Channel Auto
The Messenger Device is provided with a Multi Stage Amplification. The Primary stage of Amplification is done with help of Front End Amplifiers and Secondary stage of Amplification is done with in the Analog to digital converter (ADC) that has an Internal amplifier.
The Front End Amplifiers setting can be done using Optimize Channel Auto.
Example
1:
Adaptation
of an absorption sensor to Signal in 1:
Example 2:
Adaptation of
an scatter light turbidimeter.
The raw signal level (mV) should
normally be in a range between 1200 and 2400 mV. (2500mV out of range). In such
a case if the signal level is out of this tolerance try Optimize gain
manual first.
In case even after the manual gain
optimization one or more signals are out of tolerance, then there is every
possibility that you need to change the gain by making using of jumpers on the
hardware.
Please see the manual for details or
contact Metrisa.
In case of very low Signals (e,g.
Scatter light) measuring at a very low range the signals may even need max.
Amplification under the tolerance level.
%Range is related to the set point i.e. 100% which is 2000mv and 120% is 2500 mv which is the range of the ADC Converter
Raw mv displays the current channel mv.
Choose Channel (CH1..4)
When a Channel is selected and the Optimize Gain key is entered, the Messenger will try to set the Gain( Increase or decrease) to adjust the output level of the amplifier to 2000 mV (with range set at 100%) or higher if the 100% is higher, this relates to the A/D input of 2500 mV ( 80 to 120%).
Choose Restore Default Settings to restore the %Range to the default value i.e. 100.
The TRIMPOTS of the Front End Amplifier can be selected manually using Optimize Channel Manual.
Optimize Channel Manual
This shows the TRIMPOT settings on the front end Amplifier Range 0 to 100.
Choose Channel (1..4) and Tap/Click on "+" or "-" In order to increment or decrement the TRIMPOT Settings.
Clear/Reset Menu
Main Menu
>Diagnostic Menu > Clear/Reset Menu.
Choose System Reset to reset the Hardware.
Choose Default Values to default the parameters of the Messenger.
On Selection of Default Values, Factory calibration table will not be defaulted.
Main Menu > Setup Results Menu >Set
Results Offset.
Choose OFFSET TO RESULT (1..4)text box to specify the Offset. The offset can be a
postive or negative value.
The Offset Specified here will be a part of the Setup Results Equations.
Dark Current Auto Zero
Main Menu > Setup Inputs/Outputs > Set Messenger Unit
ID.
The UNIT ID is necessary when multi-units are connected on the same bus.
Choose Unit ID to specify the ID of the Messenger. The Unit ID should lie between 1..255.
Choose Messenger Tag to specify the extra information for easy identification when multiple devices are connected in the network.
Clear Results
Choose Clear Results (1-4) to Clear the Historic data stored in the buffer.
ADC-GAIN Manual
This is a secondary stage of Amplification setting on the ADC chip where we have a FRONT end Amplification setting using the electronic trim pots which can be adjusted using
Optimize Channel Auto
or
Optimize Channel Manual.
ADC gain setting allows a second stage of amplification when we are looking at very low signals and we need to amplify further beyond the front end range adjustment allowed.
This of course will cause a higher offset voltage and may reduce the range of the detection because of this offset potential that is due to the multi-stage amplification and the initial offset potential of the front-end op-amps.
The settings are defaulted to 1 but may be adjusted to 2,4,8,32,64,128
Choose Channel (1..4) and Tap/Click on the "+" or "-" in order to increment or decrement the gain setting.
Software Revision
Main Menu >Diagnostic Menu >
Software Revision.
The Menu displays the current version number of GUI and Firmware.
The version number
of the firmware will be displayed only when GUI is connected Online.
Calibrate Analog Outputs
Main Menu > Diagnostic Menu > Adjust Hardware Settings >
Calibrate Analog Outputs.
You can Calibrate Analog Outputs or adjust the 0/4- 20 mA loop, this is like a trim pot adjustment.
For example , Assign ANALOG OUT1 to 4-20 mA, 4mA ( Zero setting ) and 20 mA ( Factor setting) as the technician sets these loops he can iterate between the zero and factor to adjust the 4-20 mA exactly to the right levels.
Choose CH(1..4)
Choose Zero Setting and
Tap/Click on "+" or "_" in order to increment or decrement by 0.025.
Choose Factor Setting and Tap/Click on "+" or "-" In order to increment or decrement by 0.0005 value.
File Menu
Main Menu > Diagnostic Menu > File Menu.
The parameters set by the user can be saved and These can be retrieved at a later point of time.
Choose Save Parameters to File. You will see a dialog pop up Specify the name of the file in which the parameters need to be saved. These files will be stored with .pda extension and are referred as PDA Files.
Choose Load Parameters from File. You will see a dialog pop up. Select the file in which the parameters have been stored earlier. During the process of reading the data from the File and Transferring to Messenger Device, the percentage of data that has been transferred will be displayed on the button.
Choose Save Previous Data to File to read the previous data from messenger device and store in the Specified File.
Choose Load Firmware to prepare the Messenger Device ready for Installation of Firmware.
Data Storage Size
Main Menu > Setup Results Menu > Setup Data
Logging > Set Data Storage Size.
Choose on Result Storage Size (1..4)to Specify the storage size of each result.
If the data to be stored exceeds the Result Storage Size, existing data is overwritten
Choose Default Values to get back to default values (i.e. 500)
Total Maximum stored Results can't exceed 8000.
Store Channel Reference
Main Menu > Diagnostic Menu > Adjust
Hardware Settings > Set Channel Reference.
Using this option you can store a reading from a channel, which allows us to store an initial reference reading for the channel.
For example
when doing a color measurement with a single channel (detector) the equation for the sensor may look like the following: LOG REF CH1 - LOG CH1. Where LOG REF CH1 is a constant and it relates to the initial energy level with clear water in the cell.
Store Reference
Choose CH(1..4)
When a channel is selected and Store Reference key is entered the current reading of the channel will be stored as Reference.
When a channel is selected and Reset Reference key is entered the Reference value will be defaulted to zero.
Normalization
Choose CH(1..4)
Choose the Channel (CH1 / CH2 / CH3 / CH4) with which the selected channel needs to be normalized and Normalized Key is entered. The Channel gets normalized.
Normalized channel values are displayed in Green color.