Main Menu 

This menu helps in navigating to various Messenger options.

Choose Main Menu

  1. Calibration Menu
  2. Setup Results Menu
  3. Setup Inputs/Outputs
  4. Diagnostic Menu
  5. Set Cleaning Event
  6. Set Access Code
  7. Set Access Level
  8. Measurements screen

Calibration Menu

Main Menu  > Calibration Menu

This menu is used to correlate raw data collected from the sensors (photo detectors) corresponding to a specific standard or sample concentration. Up to 8 different samples can be read by the instrument. From this information the instrument can construct a lookup table.

You have two calibration tables. Factory calibration table and Process calibration table. Choose the desired calibration table, which needs to be changed by selecting the respective button.


Factory calibration table can be modified / viewed using FACTORY level Access permissions.

 Setup Inputs/Outputs Menu

You can configure the Inputs / Outputs of the Messenger Device.

Choose Main Menu > Setup Inputs / Outputs Menu. You will see the following Options.

  1. Set LCD Display
  2. Set Analog Outputs
  3. Assign Digital Inputs
  4. Set COM Port
  5. Set Alarms
  6. Set Messenger Unit ID
  7. Set Relay Functions

 

Set Date Time

Content Of the Topic...

 

Com Port

 Main Menu > Setup Inputs/Outputs > Set Com Port.

 

This menu is used to specify the Port settings. Speeds of up to 57K are allowed with an option to customize most of the standard options of RS-232/485 communications.

Choose COM Toggle Button to Specify the Com Port ( 1..4) that can be used for communication with the Messenger Device.

Choose Baud Toggle Button ( 4800 / 9600/ 19200/ 38400/ 57600 ) to Specify the Baud rate of Communication.

Select the Modbus toggle button to switch around the transmission protocol of "Modbus normal" to "Modbus slow" (communication with older computers).

Analog Outputs

Main Menu > Setup Inputs/Outputs> Set Analog Outputs.

The system has FOUR independent and assignable Analog 0/4 -20 mA loops, which can be connected to External Devices. Each output will operate from a user-defined RESULT equation. The sub-menus allow you to set all the necessary parameter values for Low and High limits and the required mode for the analog output.

Overrange

Analog Outputs can flag abnormal conditions or upset conditions that can be caused by Malfunction of Optics or by some material coating over the windows. We can assign a range of mv for normal operation. If the range exceeds, a fault is registered and an appropriate error message ( OPTICS FAULT ) will be displayed on the LCD and Measurements Screen.

How to configure Analog Outputs

Alarms

Main Menu > Setup Inputs/Outputs > Set Alarms

You can setup Alarm set conditions. Each Alarm can have a HYSTERESIS and a TIME DELAY (the condition must be met for at least this time before the alarm is reported). The alarm mode (LOW , HIGH , LOW LATCH , HIGH  LATCH) can be selected per individual Alarm.

Alarms can be assigned to the following signals

  1. Not Assigned
  2. Result1
  3. Result2
  4. Result3
  5. Result4
  6. Ambient Temp
  7. General Alarm


 The different Alarm Types are

  1. Low
  2. Low Latch
  3. High
  4. High Latch


 Low
Alarm is active when the measured value falls below the specified LIMIT.Alarm is released automatically
when the measuring value is above the specified LIMIT.
Low Latch
Alarm is active when the measured value falls below the specified LIMIT.Alarm should be released manually using RESET ALARMS menu
High.
Alarm is active when the measured value falls above the specified LIMIT.Alarm is released automatically when the measuring value is below the specified LIMIT.
High Latch
Alarm is active when the measured value falls above the specified LIMIT.Alarm should be released manually using the RESET ALARMS menu

Choose on the Alarm (1-4) .

Specify the signal to which the Alarm needs to be assigned to, by Choosing on the Assigned Signal Toggle Button.

Choose Type Toggle Button to specify the type of Alarm.

Choose  Limit text box to specify the mv. Choose Delay on Make/Break Text box to specify the delay in which the condition needs to be met at least before the alarm is reported.

Choose on Hysteresis to Specify the Hysteresis value.

The Hysteresis value will not have any effect for High and Low Latches.

In case any of the Alarms are not used, make sure that Type and Assigned Signal  are "NOT ASSIGNED".

Set Cleaning Event

Main Menu > Set Cleaning Event

You can program up to TWO cleaning events.

A cleaning event is an automatic cycle that allows the optics to be CLEANED by the introduction of a cleaning solution to flow through the cell.

For each Cleaning Event, you can specify the Cleaning Frequency , Cycle Duration  and Purge Time ( Time to allow the  Sample to flow and purge out the cleaning solution). A Relay can be assigned to a Cleaning Event  which will be active only during the Cycle Duration.

Cleaning Events can be activated in 2 ways, Automatically or through Assigned Digital Inputs .

Automatic

Choose Cleaning Event 1..2
Specify the Cleaning Frequency , Cycle Duration , and Purge Time.   The Cleaning Event gets activated.
Once you save the settings.

 Assign Digital Inputs

Choose Cleaning Event 1..2
Set Cleaning Frequency in Hours = 0
Specify the Cycle Duration and Purge Time. Save the settings. Activate the Relay through Assign Digital Inputs. The Cleaning Cycle gets activated ONLY ONCE.

Whenever you are trying to activate a Cleaning Event through Assign Digital inputs make sure that the Cleaning Frequency in Hours is 0. Failing so may cause malfunctioning of the Cleaning Event.

Set Access Code

Main Menu > Set Access Codes.

You can set personal access code for Operator , Technician and Factory Levels  preventing unauthorized personnel from altering key parameters that affect the operation of the unit.

Choose Operator text box. An Interface pops up Specify the Operator Password.

Choose  Technician text box. An Interface pops up Specify the Technician Password.

Choose  Factory text box. An Interface pops up Specify the Factory Password.

View/Edit Calibration

The selected Sensor number appears on the top. The data present in the selected calibration table  (Process/Factory) can be viewed/edited using the following option.


Samples NOT used in Calibration process must always contain a “0”.

 

Process Calibration

  1. Build Calibration
  2. View/Edit Calibration
  3. Fit Data
  4. Copy Fact Table  

 

Build Calibration

For each Sensor, up to 8 different samples can be used during the calibration process.
  1. Select the Sensor number by choosing on the Sensor Toggle Button.
  2. Fill sensor with sample of known concentration.
  3. Protect sensor against ambient light.
  4. Select the Sample number by choosing on the Sample Toggle Button.
  5. Enter the User Value ( Ex:- Known concentration value of a Standard ) in the text box provided. This value can also be changed using View/Edit Calibration menu.
  6. Choose Read New Data to capture the data.
  7. Choose  Store New Value to store the data.


Repeat steps 2 to 7 for each sample.

Make sure that the entered samples are evenly distributed over your range.
For example at a range 0-100 g/L you can use samples of approx. 0 gr/L, 30 gr/ L, 60 gr/L and 100 gr/L for calibration.

 

Diagnostic Menu

This menu is provided with diagnostic tools.


Choose Main menu > Diagnostic Menu,you will see a list of Diagnostic options.

 

  1. File menu
  2. View Channel Output
  3. Adjust Hardware settings
  4. Hardware Test
  5. Software Revision
  6. Clear Reset Menu

 


Hardware Test

This menu is used for testing of hardware.

  1. Lamp Control
  2. Test Analog Outputs
  3. Test Relays

Lamp Control

Main Menu >Diagnostic Menu > Hardware Test > Lamp Control.  

You can view the current lamp's status and Lamp Intensity.


OK in Lamp Status indicates BOTH Lamps are Working Correctly.
MAIN FAIL in Lamp Status indicates INTERNAL Lamp Failure.
EXT FAIL in Lamp Status indicates EXTERNAL Lamp Failure.
BOTH FAIL in Lamp Status indicates INTERNAL and EXTERNAL Lamp Failures.

Choose Lamp Intensity text box to specify the percentage of Intensity within the range 80% to 120%.

Choose Display Fault Toggle( INT LAMP / EXT LAMP / BOTH LAMPS ) button to explicitly generate Faults.

Choose Turn Lamp ON / Turn Lamp OFF Toggle Button to specify Whether Lamp needs to be Switched ON / OFF.

Choose External Lamp ON / External Lamp OFF Toggle button to specify whether External Lamp needs to be ON/OFF.
 External Lamps are operated with an optionally available separate power pack.

Test Analog Outputs

Main Menu >Diagnostic Menu > Hardware Test > Test Analog Outputs.

 

This Menu displays the mv of the FOUR Analog Outputs. A Fault in the Analog Output is displayed as "FAULT: 2mA" where 2mA is the current assigned during the fault condition. You can also test the Analog outputs using this option.

For Example:-

 Set Analog Out1 to the following parameters
OUTPUT = Result1
Low = 0mv, High = 100mv
Mode = 4-20mA
OVERRANGE = CH1
Low mv = 0 , High mv = 150


Choose Analog OUT1

Enter the 50 Milli volts in the Analog Output text box and choose the "Update Outputs" key. The current observed at the Analog OUT1 should be 12mA.

Choose Restore Previous  to restore the Analog Outputs to previous values before the entry of 50 Milli volts.

View Channel Output

Main Menu > Diagnostic Menu > View Channel Output.

 Here you can view the outputs from all the 4 channels and Messenger Ambient Temp.

Raw Data
Raw Data is channel mv displayed irrelevant of  ZERO DARK CURRENT.

Channel Signal
Channel signal is Channel mv displayed after subtracting  ZERO DARK CURRENT.

Choose Show Raw Data / Show Channel Signal Toggle Button to switch between the display of Raw data and Channel signal.

Adjust Hardware Settings

Using this menu. You can adjust the hardware settings.

  1. Zero Dark Current 
  2. Optimize Channel Auto
  3. Optimize Channel Manual
  4. Set Dark Current Auto Zero
  5. Set Channel Reference
  6. Calibrate Analog Outputs
  7. Set ADC-GAIN Manual

 

Optimize Signal Level

Content to be Filled.

 

Zero Dark Current

Main Menu > Diagnostic Menu > Adjust Hardware Settings > Zero Dark Current.

You can store the Zero Dark Current using this menu.


 Choose "ZERO CHANNELS" to store the dark current. The lamps are turned OFF and the system will take a reading for ALL the channels and stores it as Dark current.

Choose "Restore Defaults Settings" to set the Zero Dark Current to 0.


The dark current is typically very low and should not exceed by100 mV.
 

Set Unit ID

Content Of the Topic...

 

Select Language 
 
You can select the language which needs to be displayed. Currently four languages are available for selection.

  1. English
  2. Deutsch
  3. Francais
  4. Espanol


Choose any of the language buttons to switch to that particular language.

Set Access Level

Main Menu > Set Access Level.

You can enter the Access Code. Based on the Access code entered the Access Level changes.

Choose New Access Code text box. An Interface pops up Specify the new Access Code.

On entering incorrect access code the Access level will be defaulted to OPERATOR.

Fit Data

Choose the type of  Fit to be performed on the Calibration data. There are 4 options available.    

  1. Linear Fit Process Data
  2. Curve Fit Process Data
  3. Linear Fit Factory Data
  4. Curve Fit Factory Data


Choose Setup Sensor Equations to view/edit the Sensor Equations.


Minimum 2 samples must be read to perform Linear Fit.

Minimum 3 samples must be read to perform Curve Fit.

Equation Calculator

Content Of the Topic...

 

Setup Results

The Setup Results are used for formatting the output that is to be displayed to the user, which is captured from different sensors

Choose Main Menu > Setup Results Menu. You will see different Options to format the output

  1. Setup Results Nametags
  2. Setup Results Equations
  3. Setup Sensor Equations
  4. Setup Numerical Results
  5. Setup Graphical Results
  6. Set Results Units
  7. Setup Data logging
  8. Bubble Reject Threshold
  9. Set Results Format
  10. Set Results Offset

Results NameTag

Main Menu > Setup Results Menu > Setup Results NameTag

Choose the Result (1..4) Text box to specify the Nametags to be displayed for results captured in the Measurements Screen.

Example: - If the data captured in Result1 is Turbidity then the Nametag for Result1 can be given as "Turbidity".

 

Results Equation

Main Menu > Setup Results Menu > Setup Results Equations.

Choose the Result ( 1..4) text box to specify the Result Equations.

Rules for Result Equations

  1. An Equation can consist of 5 terms combined by operators.
  2. Each term can be a sensor or logarithm of sensor and can have a numerical factor.
  3. The fifth term is +/- offset and can only be added.


Example : -
The Equation is constructed by "term1 [op] term2[op] term3 [op] term4 + offset" where each term can be a sensor or the logarithm of sensor and has a numerical factor(i.e. 3.67LogSEN4).The operators [op] can be +,-,*,/

Example:-
If you want to subtract the reading of the first sensor (Sensor 1) from the reading of the second sensor (Sensor2).


Input the following :"SEN1-SEN2"

Graphical Results

Main Menu > Setup Results Menu >Setup Graphical Results.

You can format the Graphical display in Measurements screen. Check on Show Result checkboxes to select which of the results need to be displayed during the Graphical display.

Choose Range of Values to specify the Lower & Upper Limits of the Graph.

Choose Time Step in Minutes to specify the time duration with which the X-Axis Scale needs to be displayed.

Choose Show Color to select the color with which RESULT (1..4) need to be displayed in BAR / LINE Graph. 


Time Step in seconds can be specified by entering decimal values  Ex :- 0.1 , 0.2. 

 

Numerical Results

Main Menu > Setup Results Menu> Setup Numerical Results.

You can format the Numerical display in Measurements screen. Check on relevant Show Result checkboxes to select which results need to be displayed during the Numerical display in Measurements screen.

Choose Font Size Toggle Button to specify the Font Size (Small / Medium/ Large ) with which the results need to be displayed.

 

Data Logging

Main Menu > Setup Results Menu > Setup Data Logging.

You can specify the Time Interval (Sec) with which the data needs to be logged in buffer for each result.

Choose Store Result1 ( 1..4)  text box to specify the time in seconds.

Choose on Set Buffer Size to Specify the Buffer Size for each Result.

In order to default the values Choose Set Default Values  (i.e. 60 Seconds).

Clear Process Calibration

Main Menu >Diagnostic Menu > Clear/Reset Menu >Clear Process Calibration. 

Choose
Sensor (1..4).

On Selection of Sensor, choose Copy From Factory Table key.
Factory calibration table contents of the Selected Sensor are copied to the Process calibration table. 

On Selection of Sensor, choose Clear Table key.  Process calibration table contents of Selected Sensor are defaulted to 0.

 

Clear Factory Calibration

Main Menu >Diagnostic Menu > Clear/Reset Menu >Clear Factory Calibration.
 

Choose Sensor (1..4).

On Selection of Sensor when Clear Table key is entered. Factory calibration table contents of selected sensor are defaulted to 0.

 

 Digital Inputs

Main Menu >Setup Inputs/Outputs>Assign Digital Inputs.

The Digital Inputs are usually connected to External Devices Which may want to send a command to the Messenger Device. This Helps the Messenger Device to synchronize with other devices that are a part of the process being monitored.

Following are the commands that can be issued from the External devices to the Messenger Device.

  1. Not Assigned
  2. Activate Alarm1
  3. Activate Alarm2
  4. Activate Alarm3
  5. Activate Alarm4
  6. Activate Relay1
  7. Activate Relay2
  8. Activate Relay3
  9. Activate Relay4
  10. General Fault
  11. Activate Clean Cycle1
  12. Activate Clean Cycle2
  13. Start/Stop Result1
  14. Start/Stop Result2
  15. Start/Stop Result3
  16. Start/Stop Result4
  17. Start/Stop ALL


Choose Input (1..4)  to Assign the commands that are mentioned above to the Digital Inputs.

Choose Restore Default Settings to switch back to the Default parameters.

Relay Functions 

Main Menu > Setup Inputs/Outputs > Set Relay Functions.

This menu allows the user to assign the FUNCTION of the Relay.

The   Functions available are :   

  1. NOT ASSIGNED
  2. Alarm 1
  3. Alarm 2
  4. Alarm 3
  5. Alarm 4
  6. General Fault
  7. Cleaning device 1
  8. Cleaning device 2

Relay functions in 2 states
  1. Active ON
  2. Active OFF Fail -Safe


 

Sensor Equations

Main Menu > Setup Results Menu >Setup Sensor Equations.


Choose the Sensor  (1..4) text box to specify the Sensor Equations.

Rules for Sensor Equations

  1. An Equation can consist of 5 terms combined by operators.
  2. Each term can be a channel or logarithm. of channel and can have a numerical factor
  3. The fifth term is +/- offset and can only be added.


Rules : -
 The Equation is constructed by "term1 [op] term2[op] term3 [op] term4 + offset" where each term can be a channel or the logarithm of channel and has a numerical factor(i.e. 3.67LogCH4).The operators [op] can be +,-,*,/

Example 1:
Scatter Light Turbidity:

Example 2:
Absorbance:

Example 3:
Dual Channel Absorbance (Color):


Results Units

Main Menu > Setup Results Menu > Set Results Units.

Choose the Result  (1..4) Text box to specify the units to be displayed for results captured in the Measurements Screen.

Example: - If the data captured in Result1 is Turbidity then the unit of Result1 can be given as "NTU".
Typical Units are ppm, TEF, FTU, NTU, mg/l, g/l, %, APHA, etc.. 

 

Test Relays

Main Menu >Diagnostic Menu>Hardware Test> Test Relays.

This Menu displays the current status of the FOUR Relays i.e. Signal the Relay has been assigned and whether it is switched ON / OFF .

Choose the ON / OFF Toggle button to Switch ON / OFF the Relay.

Review/Reset Alarms

Main Menu >Diagnostic Menu > Clear/Reset Menu >Reset Alarms. 

This menu displays the status of the 4 Alarms.

An ACTIVE in the status block with RED background indicates Alarm is ON. 


In order to clear the Alarm, choose the Alarm with ACTIVE in the Status block. and choose the Clear ALARM key.

Start / End Menu

  1. Select Language
  2. Main Menu
  3. Select Messenger
  4. Go ONLINE
  5. Exit

 

Select Language

 

Select Language in which you want to view the help in

  1. English
  2. German
  3. French
  4. Spanish

 

Bubble Reject Threshold

Main Menu > Setup Results Menu > Bubble Reject Threshold.

Bubble Reject filter is used to smooth system response when bubbles are passing. and is based on a data averaging and acts upon a threshold limit set by the user in % change detection.

Choose Bubble Reject text boxes to specify the % of  Threshold limit.

Choose Filter Delay text boxes to specify the delay in sec.



Example:
Bubble reject =20%
Filter delay = 30sec
These values will affect the displayed measurement results. The Maximum change allowed is 20% in 1 sec with a settling time of 30 seconds.

Results Format

Main Menu > Setup Results Menu > Set Results Format.

You can format the results displayed in Measurements screen.

Choose Number of Decimal Places for each Result to specify the precision with which the result needs to be displayed in Measurements Screen.

Choose Allow Values for each Result to specify Whether POS( Positive values) or ALL ( Positive and Negative values ) to be displayed in Measurements Screen.

Optimize Channel  Auto

Main Menu > Diagnostic Menu >Adjust Hardware Settings>Optimize Channel  Auto.

The Messenger Device is provided with a  Multi Stage Amplification. The Primary stage of Amplification is done with help of Front End Amplifiers and  Secondary stage of Amplification is done with in the Analog to digital converter (ADC) that has an Internal amplifier.

The Front End Amplifiers setting can be done using  Optimize Channel Auto.

Example 1:
Adaptation of an absorption sensor to Signal in 1:

Example 2:
Adaptation of an scatter light turbidimeter.

The raw signal level (mV) should normally be in a range between 1200 and 2400 mV. (2500mV out of range). In such a case if the signal level is out of this tolerance try “Optimize gain manual” first.

In case even after the manual gain optimization one or more signals are out of tolerance, then there is every possibility that you need to change the gain by making using of jumpers on the hardware.

Please see the manual for details or contact Metrisa.

In case of very low Signals (e,g. Scatter light) measuring at a very low range the signals may even need max. Amplification under the tolerance level.

%Range
is related to the set point i.e. 100% which is 2000mv and 120% is 2500 mv which is the range of the ADC Converter 

Raw mv displays the current channel mv.

Choose Channel (CH1..4)

When a Channel is selected and the Optimize Gain key is entered, the Messenger will try to set the Gain( Increase or decrease) to adjust the output level of the amplifier to 2000 mV (with range set at 100%) or higher if the 100% is higher, this relates to the A/D input of 2500 mV  ( 80 to 120%).

Choose Restore Default Settings to restore the %Range to the default value i.e. 100.

The TRIMPOTS of the Front End Amplifier can be selected manually using Optimize Channel Manual.

Optimize Channel Manual

Main Menu > Diagnostic Menu >Adjust Hardware Settings>Optimize Channel  Manual.

This shows the TRIMPOT settings on the front end Amplifier Range 0 to 100.

Choose Channel (1..4)  and Tap/Click on "+" or "-" In order to increment or decrement the TRIMPOT Settings.



In case if you change the amplification, please cross verify the raw signal level (mV) in " Optimize Channel Auto". The optimum Signal level should lie between 2000 and 2200mV. For further reference please refer the help content in sub menu of Optimize Channel Auto.

Clear/Reset Menu

Main Menu >Diagnostic Menu > Clear/Reset Menu.

  1. System Reset
  2. Default Values
  3. Clear Process Calibration
  4. Clear Factory Calibration
  5. Clear Results
  6. Reset Alarms


Choose System Reset to reset the Hardware.

Choose Default Values to default the parameters of the Messenger.

 On Selection of Default Values, Factory calibration table will not be defaulted.

Results Offset

Main Menu > Setup Results Menu >Set Results Offset.

Choose  OFFSET TO RESULT (1..4)text box to specify the Offset. The offset can be a postive or negative value.

The Offset Specified here will be a part of the Setup Results Equations.

 

 Dark Current Auto Zero

Main Menu > Diagnostic Menu >Adjust Hardware Settings>Set Dark Current Auto Zero.

Choose Auto Zero Frequency in Hours text box to specify the Time Interval (Hours) at which the Dark current should be made zero. 

Messenger Unit ID

Main Menu > Setup Inputs/Outputs > Set Messenger Unit ID.

The UNIT ID is necessary when multi-units are connected on the same bus.

Choose Unit ID to specify the ID of the Messenger. The Unit ID should lie between 1..255.

Choose Messenger Tag to specify the extra information for easy identification when multiple devices are connected in the network. 

 

Clear Results

Main Menu >Diagnostic Menu > Clear/Reset Menu >Clear Results.
 

Choose Clear Results (1-4) to Clear the Historic data stored in the buffer.

 

ADC-GAIN Manual

Main Menu >Diagnostic Menu >Adjust Harware Settings>Set ADC-GAIN Manual.

 

This is a secondary stage of Amplification setting on the ADC chip where we have a FRONT end Amplification setting using the electronic trim pots which can be adjusted using 
Optimize Channel Auto 
or  
Optimize Channel Manual.

ADC gain setting allows a second stage of amplification when we are looking at very low signals and we need to amplify further beyond the front end range adjustment allowed.

This of course will cause a higher offset voltage and may reduce the range of the detection because of this offset potential that is due to the multi-stage amplification and the initial offset potential of the front-end op-amps.

The settings are defaulted to 1 but may be adjusted to 2,4,8,32,64,128

Choose Channel (1..4)  and Tap/Click on the "+" or "-"  in order to increment or decrement the gain setting.

Software Revision

Main Menu >Diagnostic Menu > Software Revision. 

The Menu displays the current version number of GUI and Firmware.

The version number of the firmware will be displayed only when GUI is connected Online.


Calibrate Analog Outputs 

Main Menu > Diagnostic Menu > Adjust Hardware Settings > Calibrate Analog Outputs.

You can Calibrate Analog Outputs or adjust the 0/4- 20 mA loop, this is like a trim pot adjustment.

For example , Assign ANALOG OUT1 to 4-20 mA, 4mA ( Zero setting ) and 20 mA ( Factor setting) as the technician sets these loops he can iterate between the zero and factor to adjust the 4-20 mA exactly to the right levels.

Choose CH(1..4)
Choose Zero Setting  and Tap/Click on "+" or "_" in order to increment or decrement by 0.025.

Choose Factor Setting and Tap/Click on "+" or "-" In order to increment or decrement by 0.0005 value.

 

File Menu

Main Menu > Diagnostic Menu > File Menu.

 The parameters set by the user can be saved and These can be retrieved at a later point of time.

Choose Save Parameters to File. You will see a dialog pop up Specify the name of the file in which the parameters need to be saved. These files will be stored with .pda extension and are referred as PDA Files. 

Choose Load Parameters from File. You will see a dialog pop up. Select the file in which the parameters have been stored earlier. During the process of reading the data from the File and Transferring to Messenger Device, the percentage of data that has been transferred will be displayed on the button.

Choose Save Previous Data to File to read the previous data from messenger device and store in the Specified File.

Choose Load Firmware to prepare the Messenger Device ready for Installation of Firmware.


This function needs additional software please contact Metrisa.

 

Data Storage Size

Main Menu > Setup Results Menu > Setup Data Logging > Set Data Storage Size.

Choose on Result Storage Size (1..4)to Specify the storage size of each result.

If the data to be stored exceeds the Result Storage Size, existing data is overwritten

Choose Default Values to get back to default values (i.e. 500)

Total Maximum stored Results can't exceed 8000.

Store Channel Reference

Main Menu > Diagnostic Menu > Adjust Hardware Settings > Set Channel Reference.

Using this option you can store a reading from a channel, which allows us to store an initial reference reading for the channel.
 
For example

when doing a color measurement with a single channel (detector) the equation for the sensor may look like the following: LOG REF CH1 - LOG CH1. Where LOG REF CH1 is a constant and it relates to the initial energy level with clear water in the cell.

Store Reference

Choose CH(1..4)
When a channel is selected and Store Reference key is entered the current reading of the channel will be stored as Reference.

When a channel is selected and Reset Reference key is entered the Reference value will be defaulted to zero.

Normalization

Choose CH(1..4)
Choose the Channel (CH1 / CH2 / CH3 / CH4) with which the selected channel needs to be normalized and Normalized Key is entered. The Channel gets normalized.

 Normalized channel values are displayed in Green color.

Introduction

The multifunctional transmitter Model Messenger can be used with all sensors of the MoniTurb / MoniSpec series. The sensors of the MoniTurb series use the principles of forward and/or side scattered light for the detection of turbidity.

The sensors of the MoniSpec series (single channel - or dual wavelength - absorption) detect solid (or alternate colour) concentrations. Depending on the sensor type, the absorption measurements can be done in the UV, VIS or NIR range.

The Messenger even allows for simultaneous use of multiple sensors. Hereby you can use up to four sensors (single channel absorption). The measurement results can be linked together using almost any mathematical equation. This allows for easy implementation of dosage systems (difference measurement).

The programming / calibration of the system can be done via a PC or a handheld computer, using the menu-based software. The integrated data logger / paperless recorder simplifies the evaluation and archiving of measurement results.

 

Introduction

 

Concepts

  1. Online/Offline
  2. Save and Read Messenger Parameters
  3. Access Levels
  4. Calibration of sensors
  5. Setting up Results
  6. Setting Inputs/outputs
  7. Diagnostic Menu
  8. Measurements

       

Online/Offline

 The Software Functions in 2 modes


  1. ONLINE
  2. OFFLINE


Modes can be switched by pressing the Online / Offline Toggle button

Each time the mode is switched, the software permits synchronization of the messenger i.e. either Sending all parameters to messenger, or Fetching all parameters from messenger

 

Messenger Client Mode(Online)

In this mode Software Permits
  1. Editing of Parameters. Parameters are saved to messenger when exiting from different parameter screens.
  2. Saving Messenger parameters.

     

File Editor Mode(Online/Offline)

In this mode Software Permits
  1. Editing Parameters
  2. Reading Messenger Parameters
  3. Saving Messenger Parameters

 

Save Messenger Parameters

Messenger parameters can be saved in 2 ways Using Write File Option File Option Present in Diagnostic Menu.

Using Write File Option

  1. Choose Write File Option in Startup menu.
  2. Specify the file in which the parameters need to be saved.

File Option Present in Diagnostic Menu.

Choose Main Menu > Diagnostic > File Menu > Save PDA Parameters to a PDA File.

Specify the File in which the parameters need to be saved.

 

Read Messenger Parameters

Choose Read File Option in Startup Menu.

Select *.mpr or *.pda Files. To read the Messenger Parameters.

 

Access Levels

The GUI menus are layered using access codes to avoid inadvertently altering key parameters by an unskilled operator.

Three access levels are available: Operator, Technician and a hidden layer for Factory.

The Operator Level is the default Level.

The Technician level requires a more skilled operator and a better understanding of what is needed especially to execute a good calibration curve.

The Factory level is accessible by specific code to allow us to change the personality of the GUI and/or Messenger system features or options.

See Also

Set Access Code

Set Access Level

 

Factory Calibration

In this submenu the standard calibration of the Messenger Device is accomplished.

  1. Build Calibration
  2. View/Edit Calibration
  3. Fit Data
  4. Copy Fact Table


This calibration possibility is available only with Factory level password

 

Measurements Screen

LCD Display

Main Menu > Setup Inputs/Outputs > Set LCD Display.

 The LCD display Contains 4 Lines which can be configured to display Results ,  Sensors and Temperature values. You can adjust the contrast of the LCD display.

Choose Line (1..4) . Enter the text  which needs to appear on the Line. Select the Signal(Results / Sensors / Ambient) that needs to appear along with the selected Precision.

In order to adjust the contrast of the LCD, Choose "+" / "-" to increment or decrement the value. The default contrast value is 120.

The Maximum text that can be displayed on a Line is 20 characters. When a Signal is assigned only 11 characters of text entered will be displayed and the remaining characters get truncated.

Select Messenger

Using this option you can search for the Messenger Devices present in the network. Messenger Devices present in the network can be identified by specifying the START and END parameters(Device ID's) and then choose SEARCH button.

During the search operation the device being searched is displayed on the SEARCH button. The Messenger Devices that are identified during the search are displayed on the screen. In order to connect to a device choose the Device ID. The Device ID gets highlighted.

If you are familiar with Device ID to which you want to connect. Tap/Click on  any of the Device ID display boxes. A numeric key pad pops up then specify the Device ID

Copy Factory Table

The selected sensor's factory calibration table is copied to process calibration table

How to Configure Analog Outputs

To configure the Analog Output, Choose the ANALOG OUT (Analog Out 1-4).

Choose Output Toggle Button ( Not Assigned / Result1 / Result2 / Result3 / Result4) to specify  the user defined RESULT equation needs to be assigned to ANALOG OUTPUT. 

Choose Mode Toggle Button to Specify the Mode (0/4-20 mA ) Loops.

Choose Low, High text boxes to specify the Low and High mv.

Choose Overrange toggle button to specify the channel (CH1..4) Assign the range of mv for normal operation by choosing the Low mv and High mv.

 In case any of the Analog Output is not used, make sure that OUTPUT , OVERRANGE parameters are "NOT ASSIGNED".

ACTIVE ON

As long as assigned function is TRUE The Relay stays in activated state.

For Example

The Alarm 1...4 are selected and thresholds are defined in the Set Alarm menu.

If we set an Alarm on Result 1 exceeding a threshold of 100 and we assign Relay #1 as an indicator of such Alarm condition, the Relay#1 is activated when this threshold is reached or exceeded  based on the Alarm mode latching or NON-latching. It would reset when the Result is back below the threshold defined.

 

Active OFF Fail Safe

As long as the Assigned signal is FALSE the Relay stays in activated state.

For Example

The Alarm 1...4 are selected and thresholds are defined in the Set Alarm menu.

If we set an Alarm on Result 1 exceeding a threshold of 100 and we assign Relay #1 as an indicator of such Alarm condition, the Relay#1 is NOT activate as long as the Alarm condition is TRUE.

Previous Data

Choose OPTIONS button    on the top right corner of the measurements screen , A popup menu appears.
Choose Previous Data.

Entire Previous Data is retrieved. Most recently logged data is displayed as first frame of graph.

 Choose Previous "<" and  Forward " >" buttons on the bottom of the graph to scroll through the Previous Data

  This operation may take some time to read the data from device and process the data. 

Choose Numerical / Graph / Bar Toggle button to switch between display formats

 

GRAPH SETTINGS

This menu is used to configure the look and feel of the Graph. You can configure.


On Selection of Auto Scaling the Y-Axis limits of graph will be set based on the Max Y and Min Y values present in the data. Otherwise the range parameters specified in the Setup Graphical display will be set. 

  X - Axis labels in the graph will be any of the following formats HH:MM:SS ,  MM:SS based on the Vertical Tic Lines selected.

HH - Hours,  MM - Minutes,  SS - Seconds.

 

Online Analysis

You can view data being captured by Messenger Device ONLINE.

The data captured from the Messenger Device can be viewed in 3 different formats.

  1. Numerical Display
  2. Graphical Display
  3. Bar Chart Display


Numerical Display and Graphical Display can be customized using the Setup Graphical Display, Setup Numerical display Menus.

Choose Numerical /  Graph  / Bar Toggle button to switch between the displays.

 

Alarm status

Status of Alarms are represented on the bottom of the Mesurements screen.

Codes displayed in Alarm Status are shown below.
1 - Alarm 1 Active
2 - Alarm 2 Active
3 - Alarm 3 Active
4 - Alarm 4 Active
CC1 - Cleaning Event1 Active
CC2 - Cleaning Event2 Active 



For Example  :  "4 CC1" string display in Alarm status indicates Alarm 4 and Cleaning Event1 are active.

How can you customize Graphical and Numerical Display ?

How can you  view value of data point on a graph area ?



Tap on the data point you want to view on the graph area, Crosshair appears displaying the X-axis and Y-axis data.

Choose Previous "<" and Forward ">" buttons to scroll through the data in the currently displayed frame.

Tap outside the graph area to remove the crosshair. 

<

Save Previous Result Data

Check on Select Result textboxes to select which results need to be saved. Choose Save, A dialog pops up.specify the file name.



 

Adaptation of an absorption sensor to Signal in 1:

  1. Fill sensor with low concentration (zero liquid) to generate the maximum Detector signal.
  2. Protect sensor against ambient light.
  3. Select text box (CH1) and click the optimize gain button.
  4. In ideal case the raw mV value should change to approx. 2000mV.
  5. In case of large deviations from the 2000 mV repeat step 4.

Adaptation of an scatter light tubidimeter:
 

  1. Fill sensor with low concentration (zero liquid) to generate the maximum Direct beam Detector signal.
  2. Protect sensor against ambient light.
  3. Select text box (CH1) and click the optimize gain button.
  4. In ideal case the raw mV value should change to approx. 2000mV.
  5. In case of large deviations from the 2000 mV repeat clicking the optimize gain.
  6. Fill sensor with a calibration sample with maximum turbidity, so that it generate the maximum scatter beam Detector signal.
  7. Protect sensor against ambient light.
  8. Select text box (CH2) and click optimize gain button.
  9. In ideal case the raw mV value should change to approx. 2000mV.
  10. In case of large deviations from the 2000 mV repeat clicking the optimize gain button.

 


Scatter Light Turbidity:

For a scatter light turbidimeter the signals of DB detector and SB detector has to be equated.

Typically you can connect the DB detector to signal input of CH1 and the SB detectors to signal input of CH2 (12°) and signal input to CH3 (90°).

The equation can given as follows: SB/DB

In case if sensor 1 is defined as 12° system tap/click on text box of sensor 1” and enter the equation “CH2/CH1”.

In case if sensor 2 is defined as 90° system tap/click on text box of “sensor 2” and enter the equation “CH3/CH1”.   Absorbance:

If the detector of a single channel absorbance measurement is connected to signal input 1 (CH1 ) and defined as sensor 2, tap/click on text box of sensor 2 and enter the following equation:
"LogREFCH1 - LogCH1 " Dual Channel Absorbance (Color):

A color measurement uses two detectors, reference detector (detection of particles) and a measurement detector (detection of color and particles). The measurement detector is connected to signal in 1 (CH1). The reference detector is connected to signal in 2 (CH2 ). In case if sensor 1 is defined as color system tap/click on text box “sensor 1” and enter the equation "LogCH2 – LogCH1 ".

The measurement channel (color + particles) signal will be subtracted from the reference channel and the result will be a clean color signal.